Procedures <Prev Next>

Contour Fitting

Contour fitting allows fitting the overall shape of an experimental spectrum to a simulation, and is useful for low resolution spectra in which the rotational structure is not fully resolved. Note that the results from this sort of fit are typically less reliable than for a line position fit, and a line position fit is to be preferred if at all possible.
  1. To start a contour fit, Overlay the experimental spectrum over the PGOPHER simulation. The initial match must be quite good for a contour fit to work - a contour fit will typically not move a peak more than the peak width, for example.
  2. It is easier to work if the intensity units of both the overall overlay, and the specific overlay(s) to be floated are set to "Normalised", rather than arbitrary. Go to the overlays window (View, Overlays), click on the topmost item and ensure that "IntensityUnits" are set to "Normalized". Click on the specific overlay and check the same setting. After doing this the "Baseline" and "Scale" will need to be adjusted for a reasonable match with the simulated spectra as the vertical autoscaling is now turned off.
  3. Open the constants window (View, Constants) and click on “Fix all” to ensure that no molecular constants are floated to start with.
  4. The first parameters to be floated must be the scale and baseline of the overlayed spectrum. Go to the overlays window (View, Overlays), select the overlay required (probably the first one unless you have more than one experimental spectrum overlaid) and float these two parameters (double click in the float column so it reads “yes”). 
  5. “View”, “Fit log”, and choose “Contour Fit”. A separate plot showing the obs-calc (difference between the experimental spectrum and the simulation) is shown in the main window (see below for an example), along with the simulation. The “Log” window now shows new values for the floated parameters and their errors, the average error, and a correlation matrix. The error will typically be very large at this point
  6. Press “Contour Fit” repeatedly, until the average error is as low as possible, and a simulation that roughly matches the experimental spectrum is shown. If you have problems you may have to fix some of the parameters and try manually changing them.
  7. To further improve the fit, start floating molecular parameters. Floating many parameters simultaneously will often cause the fit to make wild changes in some of the parameters unless the fit is very good. For best results float additional parameters one or two at a time, starting with the most important. The Undo Fit button allows you go back to previous fits, if you find you have gone too far from a reasonable fit. For example, the excited state origin might be the first molecular constant to float (“View”, “Constants”, and double click on the “no” next to the constant to change it to yes.). Again, press “Contour fit” repeatedly until the error reaches a minimum, and converges.
  8. Other parameters to consider are A or B rotational constants, the linewidth and rotational temperature. Eventually, the simulation should closely resemble the experimental spectrum, and the obs-calc will be much smaller.